| Model Number |
15 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and was able to duplicate the reported issue.Physio determined that the cause of the reported issue was due to the therapy cable.Physio performed testing using a known good therapy cable and the issue was resolved.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The customer will purchase a new therapy cable and replace it.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that during a patient event, the device would not detect that electrodes were connected.The device gave a "connect electrodes" message even though the patient was connected to the device.The patient survived the event and there was no adverse effects to the patient as a result of the reported issue.No further details about the patient or the event were provided.
|
| |
|