| Model Number |
MZ9271 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer¿s report of the set leaking near the filter was confirmed.Visual inspection showed no obvious damage or issues.Pressure testing resulted in a leak at the engagement between the filter outlet and tubing.Inspection under magnification showed insufficient/lack of solvent at the tubing engagement.The root cause of the insufficient solvent is improper assembly due to equipment and/or operator error.
|
| |
|
Event or Problem Description
|
|
The set was received as an unexpected product returned.A note attached to the returned product states "leaking near filter." blood was seen in the fluid path.There was no report of patient harm.
|
| |
|