BECTON DICKINSON MEDICAL SYSTEMS 5ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML; PREFILLED HEPARIN FLUSH SYRINGE Back to Search Results
Catalog Number 306513
Health Effect - Clinical Codes Bacterial Infection (1735); Death (1802)
Date of Event 06/28/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Date of event: unknown.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that between march and april of 2018, a patient received multiple heparin flushes with a 5ml bd posiflush¿ heparin lock flush syringe, in 10ml syringe, 100 usp units/ml.In early (b)(6), the patient passed away.There was no report of infection or additional information regarding this patient's death.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and (b)(6) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and (b)(6) identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.(b)(6)'s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and (b)(6).
 
Additional Manufacturer Narrative
Additional information: please note that some of the aeg questions have been updated due to a response from the customer.Cause of death was determined to be pneumonia, heart failure.No medical intervention or other actions were taken as a result.Course of treatment was not changed.Culture was taken (b)(6) with no growth after five days.Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: death.Device single use?: no.Device returned to manufacture: no.Investigation summary: as both a lot number and a sample were unavailable for this incident, a full investigation consisting of a sample analysis and a device history record review could not be completed.Prior to april 2018, there was no trend for infection or death regarding the bd franklin products.Upon the increase of infection reports beginning in april 2018, a corrective and preventive action plan was initiated to address the issue.A review of all lot sterility testing performed for product released between april 2015 and may 2018 confirmed that no organism growth was identified for any lots released during the shelf life of this product.A direct causation between the reported infections and deaths and the bd franklin product has not been identified.
 
Event or Problem Description
It was reported that between (b)(6) and (b)(6) of 2018, a patient received multiple heparin flushes with a 5ml bd posiflush¿ heparin lock flush syringe, in 10ml syringe, 100 usp units/ml.In early (b)(6), the patient passed away.There was no report of infection or additional information regarding this patient's death.Out of an abundance of caution and in the interest of public health, bd voluntarily recalled certain lots of bd posiflush¿ heparin lock flush and bd¿ pre-filled normal saline flush syringes due to a potential for contamination with serratia marcescens bacterium.Bd was notified by the u.S.Food and drug administration (fda) and centers for disease control and prevention (cdc) about a potential epidemiological link between catheter related blood stream infections and the s.Marcescens bacterium.Specifically, the fda and cdc identified a potential connection between reports of infection in a small number of patients caused by s.Marcescens across multiple states.Cdc¿s initial investigation found that affected patients had received treatment using certain bd flush products.To date, there is no evidence of bd flush product testing positive for this bacterium.Investigations are ongoing by bd, fda, and cdc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5ML BD POSIFLUSH¿ HEPARIN LOCK FLUSH SYRINGE, IN 10ML SYRINGE, 100 USP UNITS/ML
Common Device Name
PREFILLED HEPARIN FLUSH SYRINGE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
9630 south 54th street
franklin WI 53132
MDR Report Key7696884
Report Number2134319-2018-00092
Device Sequence Number11011739
Product Code FOZ
UDI-Device Identifier30382903065135
UDI-Public30382903065135
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K011967
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue Number306513
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer06/28/2018
Initial Report FDA Received Date07/17/2018
Supplement Report FDA Received Date09/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-