| Model Number |
MMT-523LNAS |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had a compromised force sensor system alarm.Customer also reported that the device was squirting out insulin during manual prime process and the drive support cap was flush.Blood glucose level at the time of the incident was not provided.The customer was advised that the insulin pump would be replaced and agreed to return the product for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the displacement test.The device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime or compromised force sensor system test, excessive no delivery test and occlusion test or verify prime/fill anomalies due to motor error alarm.Insulin pump received with lose drive support disk.
|
| |
|