• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-523LNAS
Medical Device Problem Code Priming Problem (4040)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump had a compromised force sensor system alarm.Customer also reported that the device was squirting out insulin during manual prime process and the drive support cap was flush.Blood glucose level at the time of the incident was not provided.The customer was advised that the insulin pump would be replaced and agreed to return the product for analysis.
 
Additional Manufacturer Narrative
Insulin pump passed the displacement test.The device received with motor error alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime or compromised force sensor system test, excessive no delivery test and occlusion test or verify prime/fill anomalies due to motor error alarm.Insulin pump received with lose drive support disk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7697098
Report Number3004209178-2018-93002
Device Sequence Number1553936
Product Code OYC
UDI-Device Identifier00643169479630
UDI-Public(01)00643169479630
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA3523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2018
Initial Date Received by Manufacturer 06/18/2018
Supplement Date Received by Manufacturer09/12/2018
Initial Report FDA Received Date07/17/2018
Supplement Report FDA Received Date10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
Patient Weight148
-
-