SMITH & NEPHEW, INC. EXT SLV REMVL TOOL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
116608 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
05/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, when trying to remove sleeve from stem, the extractor got stuck in the sleeve and the sleeve was stuck in the stem.Therefore, the stem needed to be removed with the other two devices stuck.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device was returned and evaluated.A visual examination of the returned sleeve removal tool indicated that the threaded end has fractured off.A clinical analysis indicated that based on the information provided, the root cause of the acetabular protrusion and subsequent conversion to left tha is likely disease progression.The inability to remove the sleeve from the well-fixed anthology stem via alternative means as well as the re-use removal tool which fractured were the contributing factors to the stem removal.The patient impact beyond the reported clinical symptoms, surgical extension, and l hip revision/conversion to tha with an extensive osteotomy and intraoperative complications cannot be determined.No further medical assessment is warranted at this time.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate further as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|