• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-RAD MEDICAL DIAGNOSTICS GMBH BLOOD GROUPING REAGENT ANTI-FYA (FY1); SERACLONE ANTI-FYA (FY1)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-RAD MEDICAL DIAGNOSTICS GMBH BLOOD GROUPING REAGENT ANTI-FYA (FY1); SERACLONE ANTI-FYA (FY1) Back to Search Results
Catalog Number 808188100
Health Effect - Clinical Code No Information (3190)
Date of Event 06/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This is our combined initial and final report on this incident.
 
Event or Problem Description
The customer reported false negative fy(a) antigen typing results when using seraclone anti-fy(a).The customer used various fy(a+b+) positive reagent red blood cells from two different manufacturers.The customer did neither return the supposedly defective product for investigational testing nor the red blood cells that had caused the false negative test result.Therefore our quality control laboratory tested their retention sample with different donor and reagent red blood cells for potency and specificity.For the specificity testing our quality control laboratory used screening and panel cells of a competitor as well as the reagent red blood cells of the biotest cell product line.All positive and negative reactions were correct.We did not observe any false negative reaction.The required minimum titer of 16 was exceeded.Testing by our quality control laboratory confirmed that the allegedly defective lot of seraclone anti-fy(a) functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD GROUPING REAGENT ANTI-FYA (FY1)
Common Device Name
SERACLONE ANTI-FYA (FY1)
Manufacturer (Section D)
BIO-RAD MEDICAL DIAGNOSTICS GMBH
industriestrasse 1
dreieich, hessen 63303
GM  63303
MDR Report Key7697405
Report Number9610824-2018-00042
Device Sequence Number906791
Product Code KSZ
UDI-Device Identifier07611969951833
UDI-Public(01)07611969951833(17)180829(10)8636050-01
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/29/2018
Device Catalogue Number808188100
Device Lot Number8636050-01
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-