ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Moisture Damage (1405); Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a button/keypad (tactile changes w/moisture) issue.It was reported that the up, down and ok buttons were under responsive to user presses.There is no indication that the product issue caused or contributed to an adverse event.The complaint is being reported because the issue has the ability to result in inadvertent or incorrect insulin delivery or the inability to use the pump.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 23-aug-2018 with the following findings: the keypad was found to be intact; no damage was observed.All keypad buttons were found to respond appropriately during testing.The complaint of a keypad response issue was not duplicated during investigation.The keypad cover was removed and no contamination was found under the key contacts.Investigation revealed corrosion in the battery compartment.A leak test verified leaks due to a battery compartment crack torn audio bolus button cover.The pump cover was removed and internal moisture was found on the printed circuit board.Investigation revealed corrosion in the battery compartment.A leak test verified leaks due to a battery compartment crack torn audio bolus button cover.The complaint of under responsive keypad buttons was not duplicated during investigation, however, moisture was confirmed.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7697706
Report Number2531779-2018-13056
Device Sequence Number19112771
Product Code OYC
UDI-Device Identifier10840406100266
UDI-Public10840406100266
Combination Product (Y/N)Y
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Was the Report Sent to FDA? Yes
Device Age24 MO
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer06/22/2018
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date09/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age23 YR
-
-