• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; POWERED WHEELCHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; POWERED WHEELCHAIR Back to Search Results
Model Number JAZZY AIR
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 04/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
 
Event or Problem Description
Alleges coming into apartment and used leg to hold door.While holding door open, used arm to hit the joystick and it lurched fast.Did not have seat belt on.Customer fell on knee and broke her leg.
 
Additional Manufacturer Narrative
The provider and pride made several attempts to contact customer but were unsuccessful.
 
Event or Problem Description
Alleges coming into apartment and used leg to hold door.While holding door open, used arm to hit the joystick and it lurched fast.Did not have seat belt on.Customer fell on knee and broke her leg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
POWERED WHEELCHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
MDR Report Key7697839
Report Number2530130-2018-00093
Device Sequence Number996016
Product Code ITI
UDI-Device Identifier00606509100069
UDI-Public00606509100069
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K143383
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberJAZZY AIR
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer07/16/2018
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age84 YR
-
-