Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Moisture Damage (1405)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned to animas for evaluation.An evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging a power (moisture ingress) issue.The reporter alleged moisture was in the battery compartment.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Additional Manufacturer Narrative
Follow-up #1: device evaluation: the device has been returned and evaluated by product analysis on 23-jul-2018 with the following findings: during investigation, the battery compartment was cracked from the threads down to the case seal on the side and below the grip pad.The returned battery cap was undamaged and able to fit.Evidence of moisture corrosion was observed in the battery canister and on the display screen inside the pump.The leak test failed due a battery compartment leak.The pump was unable to power up and was completely unresponsive.The reported power complaint was duplicated.The pump cover was removed to find further moisture corrosion on the power printed circuit board and to the continuous glucose monitoring module.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7698525
Report Number2531779-2018-13061
Device Sequence Number19270876
Product Code OYC
UDI-Device Identifier10840406100334
UDI-Public10840406100334
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Was the Report Sent to FDA? Yes
Device Age15 MO
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer06/28/2018
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date08/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age11 YR
-
-