| Model Number |
6996SQ41 |
| Medical Device Problem Codes |
Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
| Date of Event |
07/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of medical devices: 693565 lead implanted: (b)(6) 2017; dtbb1d1 crt-d implanted: (b)(6) 2016.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead alerted for a high and undefined superior vena cava (svc) defibrillation imepdance.An svc coil fracture was suspected.The lead was reprogrammed and remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the right ventricular (rv) lead triggered the alert for high/undefined svc impedance.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|