MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPTA CRTP; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
|
Back to Search Results |
|
Model Number W1TR01 |
Device Problems
Nonstandard Device (1420); Application Program Problem (2880); Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/25/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: 429688 lead, implanted: (b)(6) 2018, 5076-58 lead, implanted: (b)(6) 2018.This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that power on reset (por) was seen on remote transmission.The device was reprogrammed and remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a software error.Analysis of the device memory indicated that the atrial and ventricular rate histogram data were missing/invalid.Analysis of the device memory indicated the battery measurement was not available.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Corrected information.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|