| Catalog Number |
366450 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® k3 edta (k3e) 12mg blood collection tubes had dark clumps in the tube which are possibly foreign material.No report of injury or medical intervention.No report of blood exposure to mucous membrane.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for foreign matter with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for foreign matter with the incident lot was observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® k3 edta (k3e) 12mg blood collection tubes had dark clumps in the tube which are possibly foreign material.No report of injury or medical intervention.No report of blood exposure to mucous membrane.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® k3 edta (k3e) 12mg blood collection tubes had dark clumps in the tube which are possibly foreign material.No report of injury or medical intervention.No report of blood exposure to mucous membrane.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacturer: yes.
|
| |
|
Search Alerts/Recalls
|