Product complaint # (b)(4).Date sent to the fda: 04/30/2019.The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Corrected information: investigation summary: it was reported barb non-engagement in tissue. twenty unopened samples of product code sxmp1b408, lot llh607 were returned for analysis. during the visual inspection of thirteen unopened samples, no defects were found on the packages.The samples were opened and the swage and attachment area of the needles were noted to be as expected.The sutures were inspected along of the strand and no defects were found.Also, ten barbs were measured, and all barbs met with the requirements.In addition, the loops were as expected.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.According to the samples condition, no barb non-engagement in tissue was found.Representative samples returned to support investigation of 3 device issues captured in reports 2210968-2018-74482, 2210968-2018-74484, 2210968-2018-74485, analysis results have been included in follow-up reports for each.
|