MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK AXS CATALYST 5 - 058X115 (FDA); CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK AXS CATALYST 5 - 058X115 (FDA); CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number M003IC0581150
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.Visual and microscopic examination of the returned device found that the dac shaft was broken/ fractured at the junction of the grilamid sleeve and the catheter shaft at the proximal end.There were no anomalies noted to the hoop and retaining clip.As per the additional information, there was no damage noted to the packaging.It is probable that the device was damaged as a result of handling of the device during removal of the device from its packaging.
 
Event or Problem Description
The review of the investigation revealed that the distal access catheter (subject device) shaft was broken.There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXS CATALYST 5 - 058X115 (FDA)
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7699905
Report Number3008881809-2018-00309
Device Sequence Number1618986
Product Code DQY
UDI-Device Identifier07613327012019
UDI-Public07613327012019
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K151667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue NumberM003IC0581150
Device Lot Number19449133
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-