Catalog Number PXX52
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.Attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The single complaint was reported with multiple events.There are no additional details regarding the additional patient events.Attempts were made to obtain the following information but no further information was available.If further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that the patient underwent a heart valve replacement procedure on an unknown date and suture was used.During the procedure, the suture was passing through tissue when the needle popped off of the suture at the swage.A like device from an alternate lot was used to complete the procedure.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Investigation summary
=
> an open box with unopened samples were returned for analysis.The product code is multi-strands double armed and must contains ten strands per packet.The issue sample was not returned for evaluation.During the visual inspection of unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area of the needles were as expected.The sutures were dispensed without problems and examined along of the strands and no defects were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the samples received, no attachment defects were found, and the tested samples met the finished goods requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHIBOND EXCEL SUTURE 30"(75CM) 2-0 G/W
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
MDR Report Key7699922
Report Number2210968-2018-74496
Device Sequence Number1108181
Product Code GAT
UDI-Device Identifier10705031049338
UDI-Public10705031049338
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Catalogue NumberPXX52
Device Lot NumberMC5180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2018
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer07/25/2018
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date08/21/2018
Patient Sequence Number1
-
-