| Catalog Number |
PLE45A |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/21/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this batch/lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date, a supplemental medwatch will be sent.To clarify, what is meant by "the routine procedure was done, the reverse was done"? did the device lock-out (no staples deployed and no cut line started)? or, did the device start to deploy staples and cut but could not be completed (partially fire)? or, did the device fully fire and stop during the automatic knife return? was the device locked on tissue with the jaws closed? if yes, how was the device removed? did the jaws eventually open? if yes, how?.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure (video gastroplasty), there was a problem with the stapler.In the last anastomosis, it presented a defect.The routine procedure was done, the reverse was done, but the device did not shoot any more staples, and did not open.It was requested a new material: 01 stapler and 01 reload.The patient had no serious adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4), date sent: 8/22/2018, batch number = > p57569.The analysis found that one ple45a device was returned with no apparent damage and with an ecr45w partially fired 1/10 cartridge loaded on the device.The device was tested for functionality in the straight position with a test cartridge reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples were noted to have the proper b-form shape.It is possible that while loading the reload, the cartridge was pushed farther back than the cartridge alignment stop windows resulting in the knife pushing the one piece sled forward and locking the cartridge.The device opened and closed without any difficulties noted.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|
| |
|
Search Alerts/Recalls
|
|
|