Model Number MMT-523NAL
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Hyperglycemia (1905); Elevated ketones/Diabetic Ketoacidosis (2364)
Date of Event 06/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that they experienced high blood glucose and diabetic ketoacidosis.The customer¿s blood glucose level was above 600 mg/dl.The customer was uncomfortable with the insulin pump.The customer declined troubleshoot for high blood glucose.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Hospitalization details has been provided which was not included with the initial report.The information has been provided with this report.(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
It was reported that the customer was hospitalized due to high blood glucose on (b)(6) 2017 with the blood glucose of above 600 mg/dl.
 
Additional Manufacturer Narrative
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had cracked reservoir tube lip and a missing end cap sticker.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7700253
Report Number3004209178-2018-93422
Device Sequence Number16420943
Product Code OYC
UDI-Device Identifier00643169693746
UDI-Public(01)00643169693746
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-523NAL
Device Catalogue NumberMMT-523NAL
Device Lot NumberB5523NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Initial Date Received by Manufacturer 06/20/2018
Supplement Date Received by Manufacturer11/14/2019
11/19/2019
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date11/16/2019
11/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age35 YR
Patient Weight140
-
-