| Model Number |
MMT-523NAL |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Elevated ketones/Diabetic Ketoacidosis (2364)
|
| Date of Event |
06/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose and diabetic ketoacidosis.The customer¿s blood glucose level was above 600 mg/dl.The customer was uncomfortable with the insulin pump.The customer declined troubleshoot for high blood glucose.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Hospitalization details has been provided which was not included with the initial report.The information has been provided with this report.(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer was hospitalized due to high blood glucose on (b)(6) 2017 with the blood glucose of above 600 mg/dl.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had cracked reservoir tube lip and a missing end cap sticker.(b)(4).
|
| |
|