If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate the complaint device was received and inspected.Visually, it was observed that the ratchet was not functional, and the jaws no longer actuated.Upon further investigation it was found that the shear pin inside the handle of the device had failed, eliminating the linkage between the jaw lever and the actuator, preventing functionality of the jaw.This shear pin failure has been typically observed in the event that excessive tissue is clamped, the linkage failure will occur within the handle and prevent loose bodies from leaving the device.This device is more than an year old and appears worn and laser markings are faded indicating heavy use.A batch record review has been conducted and the results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaint for this lot of devices that were released to distribution.At this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
It was reported by the affiliate in (b)(6) that the clever hook left device was damaged (bent/broken).It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|