• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US CLEVER HOOK LEFT; SKIN/TISSUE HOOK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US CLEVER HOOK LEFT; SKIN/TISSUE HOOK Back to Search Results
Catalog Number 214641
Medical Device Problem Codes Break (1069); Defective Device (2588); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/01/2013
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate the complaint device was received and inspected.Visually, it was observed that the ratchet was not functional, and the jaws no longer actuated.Upon further investigation it was found that the shear pin inside the handle of the device had failed, eliminating the linkage between the jaw lever and the actuator, preventing functionality of the jaw.This shear pin failure has been typically observed in the event that excessive tissue is clamped, the linkage failure will occur within the handle and prevent loose bodies from leaving the device.This device is more than an year old and appears worn and laser markings are faded indicating heavy use.A batch record review has been conducted and the results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaint for this lot of devices that were released to distribution.At this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event or Problem Description
It was reported by the affiliate in (b)(6) that the clever hook left device was damaged (bent/broken).It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEVER HOOK LEFT
Common Device Name
SKIN/TISSUE HOOK
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7700257
Report Number1221934-2018-51566
Device Sequence Number19112555
Product Code NBH
UDI-Device Identifier10886705004225
UDI-Public10886705004225
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/25/2013
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number214641
Device Lot Number13F02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2014
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2013
Initial Report FDA Received Date07/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-