| Model Number |
4244020W |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A device history record review was performed and showed that these lots of products met all requirements per the applicable manufacturing quality plan.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
During a percutaneous transluminal renal angioplasty, it was reported that the aviator plus balloon catheter was delivered to the lesion; however, it ruptured around its nominal pressure during initial inflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received via email and (description of event) was updated in the first paragraph below.During a percutaneous transluminal renal angioplasty, it was reported that the 4x20mm 142cm.014 aviator plus balloon catheter was delivered to the lesion; however, it ruptured around its nominal pressure during initial inflation.No other information was provided.There was no difficulty removing the product from the hoop.The product was removed intact from the patient.Additional procedural details were requested but are unknown.The device was not returned for analysis.A device history record (dhr) review of lot 17712403 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (although not provided) and procedural/handling factors may have contributed to the reported event since calcified/resistant lesions can damage the balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Do not exceed the rated burst pressure recommended on the label.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
During a percutaneous transluminal renal angioplasty, it was reported that the 4x20mm 142cm.014 aviator plus balloon catheter was delivered to the lesion; however, it ruptured around its nominal pressure during initial inflation.No other information was provided.The device was not returned for analysis.A device history record (dhr) review of lot 17712403 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (although not provided) and procedural/handling factors may have contributed to the reported event since calcified/resistant lesions can damage the balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Do not exceed the rated burst pressure recommended on the label.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|