• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RLT231216
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 05/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.Images were sent to gore imaging and are being evaluated.
 
Event or Problem Description
On (b)(6) 2014, the patient was implanted with a gore® excluder® aaa endoprostheses to treat an abdominal aortic aneurysm.The patient tolerated the procedure.It was reported computed tomography images (dated (b)(6) 2018), revealed there is aneurysm sac enlargement, the amount is unknown.The cause of the enlargement is unknown.The event is ongoing.
 
Additional Manufacturer Narrative
Method code 3331 addition h6: code 22 addition images were sent to gore imaging services for evaluation.Per the evaluation three time-points are available for evaluation; post-implantation cta dated january 30, 2018, post-implantation cta dated may 18, 2018, and angiogram runs dated june 21, 2018.Images dated january 30, 2018, appear to illustrate contrast pooling in the central aneurysmal sac.A lumber artery near the proximal implanted device and a set of lumbar arteries more distal appear to have contrast in them at the level of the implanted device on the january 30, 2018 and may 18, 2018, time-points.Cannot confirm communication between the lumbar arteries and contrast pooling in the sac on either time point.Comparison axial images appear to show aneurysmal growth between the first two time-points; january 30, 2018, approximately ~70mm x 70mm and may 18, 2018, approximately ~80mm x 85mm.Unable to confirm contrast outside the implanted devices in the aneurysmal sac on the angiogram imaging of june 21, 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7700581
Report Number2017233-2018-00382
Device Sequence Number13678961
Product Code MIH
UDI-Device Identifier00733132622344
UDI-Public00733132622344
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Catalogue NumberRLT231216
Device Lot Number12984237
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date07/18/2018
Supplement Report FDA Received Date08/14/2018
09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age76 YR
-
-