| Model Number |
2200-0500 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Continued from concomitant products: 250 ml exactamix baxter infusion bag; td unk.The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing leaked while connected to a patient.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The report of leaking tubing was not confirmed or replicated.Visual inspection showed a thick grey residue on the exterior of the connection between the filtered set and the bi-fuse set, indicating a potential leak site.This same residue was present on the exterior of the male luer barb of the bi-fuse set.Functional testing showed no leaking or other anomalies.A root cause for the reported failure was not determined.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tubing leaked while connected to a patient.The iv bag received with the tubing was labeled to contain 194.4 ml of tpn to be infused at 8.1 ml/hr.There was no report of patient harm.
|
| |
|