Catalog Number RMT281214
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/18/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.Per the gore® excluder® aaa endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to endoleak and aneurysm enlargement.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2013, the patient presented with an abdominal aortic aneurysm that was treated with the implantation of gore® excluder® aaa endoprostheses.On (b)(6) 2013, post-operative imaging showed an aneurysm diameter of 86mm that reportedly decreased to 81mm during follow-up imaging on (b)(6) 2016.On (b)(6) 2017, a control angiogram showed aneurysm enlargement to 83mm without evidence of an endoleak.On (b)(6) 2018, additional aneurysm enlargement to 85mm and a proximal type i endoleak were reported on a control angiogram.It was stated that the endoleak will be treated with the implantation of an aortic extender endoprosthesis on (b)(6) 2018.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7701598
Report Number2017233-2018-00380
Device Sequence Number10912698
Product Code MIH
UDI-Device Identifier00733132610679
UDI-Public00733132610679
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source study
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2015
Device Catalogue NumberRMT281214
Device Lot Number10884824
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient Weight68
-
-