MAUDE Adverse Event Report: STRYKER SPINE-US AVIATOR ASSY THREE LEVEL PLATE SIZE 48; SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US AVIATOR ASSY THREE LEVEL PLATE SIZE 48; SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS Back to Search Results
Catalog Number 48811348
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348); No Code Available (3191)
Date of Event 07/02/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent revision surgery to replace a plate and screws, as the screws were backing out.
 
Event or Problem Description
It was reported that the patient underwent revision surgery to replace a plate and screws, as the screws were backing out.
 
Additional Manufacturer Narrative
Catalog #: 48811348, lot #: 164097.Gtin #: ( b)(4).Method: visual inspection, device history review, complaint history review, risk assessment; result: the customer reported event was confirmed by the provided x-ray.A revision surgery occurred and the implants were returned for examination.The visual inspection found the plate to have a broken spring bar and screw shave no anomalies associated.Manufacturing files were reviewed and no anomalies were found.Conclusion: the exact root cause could not determined.The potential root causes include, poor surgical candidate or bone quality, excessive post-op activity by patient (not recommended by surgeon), patient does not follow surgeon instructions, patient over exerts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR ASSY THREE LEVEL PLATE SIZE 48
Common Device Name
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7702659
Report Number0009617544-2018-00177
Device Sequence Number2771467
Product Code KWQ
UDI-Device Identifier07613252152606
UDI-Public07613252152606
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48811348
Device Lot Number164097
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2018
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer09/27/2018
Initial Report FDA Received Date07/19/2018
Supplement Report FDA Received Date10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-