• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS CD HORIZON SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Catalog Number 7540220
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 06/15/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.X-ray review: in the post-op x-rays for t2-pelvis fixation no obvious hardware malfunction is identified.The patient by report is quadriplegia and there is very little soft tissue.The left side rod appears to be long, otherwise hardware placement appears appropriate.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient presented with neuromuscular scoliosis and underwent construct fixation at t2-pelvis.Reportedly post-op, the set screw failed and patient had psuedoarthrosis.On (b)(6) 2018 the patient underwent revision surgery in which the construct was explanted.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
MDR Report Key7703477
Report Number1030489-2018-01026
Device Sequence Number16332866
Product Code KWP
UDI-Device Identifier00885074468174
UDI-Public00885074468174
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K052187
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number7540220
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/19/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date07/19/2018
Supplement Report FDA Received Date10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age15 YR
Patient Weight36
-
-