| Catalog Number |
03.613.004 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information is unknown.It is unknown when device broke.Lot number is unknown.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter email address is unavailable.Reporter is a synthes employee.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that for an unknown procedure on (b)(6) 2018, the surgical team was pulling instruments and when they put the inner shaft into the removal driver they noticed the tip of the removal driver was missing as if it had been broken off (of the inner shaft).Then during the procedure, they noticed the that tip of the awl was slightly bent and very worn.It was unknown if there was a surgical delay.Patient outcome was not reported.This report is for an inner shaft for extraction screwdriver.This is report 1 of 1 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narrative: device history record (dhr) review: part number: 03.613.004; synthes lot number: 7739152; supplier lot number: n/a; release to warehouse date: (b)(4) 2014; expiration date: n/a; supplier: tecomet (b)(4).No nonconformance records (ncrs) were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.The material was reviewed and the hardness value was confirmed to meet the specification with no (relevant) non-conformance noted.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Investigation summary: customer quality investigation: the following investigations were performed: broken: device condition: visual inspection performed at customer quality (cq) on the given device showed that the inner shaft for the extraction screwdriver was broken at its tip portion that engages to the fixation pins/screw heads.The broken tip portions were not returned at customer quality.No issues were observed on other portions of the device.The given complaint condition is consistent with the device returned and therefore the complaint condition was confirmed.Dimensional analysis and document/specification review: relevant drawings for the returned instrument were reviewed (both from the time of manufacture and present revision) and no design issues were identified.Dimensional inspection performed at the device diameter at the broken region measured the dimeter as 2mm (ca-814) that falls within the specification of 2 mm +/- 0.1 mm and tolerance specified per respective inspection sheets.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.No ncrs were generated during production.Material review performed during dhr review showed no issues with device material and tis properties.Furthermore, with the given age of device (over 3 years) there is no indication that the device material and tis properties could have contributed to the complaint condition.Therefore, no design and manufacture issues were identified that would contribute to complaint condition.Conclusion: a definitive root cause for the device breakage could not be determined from the provided information.However, it is likely that any unintended excessive forces during usage/handling could have contributed to this complaint condition.The overall complaint condition is confirmed; however, no product design issues or manufacturing discrepancies were identified during this investigation based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Corrected data: lot number was not submitted in the earlier mw report in error.Date device was returned to manufacturer corrected to (b)(6) 2018.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The procedure occurred on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Lot number is now known.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|