| Model Number |
1020020 |
| Medical Device Problem Codes |
No Audible Alarm (1019); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
Patient Problem/Medical Problem (2688)
|
| Date of Event |
07/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The manufacturer received information alleging an everflo oxygen concentrator had no flow and did not alarm.The patient went to the hospital for one night.The device was evaluated by a third party service center.The customer's complaint was not confirmed.The device was found to operate and alarm to specifications.The device was found to be contaminated.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported the awareness date was (b)(6) 2018.The correct awareness date is (b)(6) 2018.
|
| |
|