• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX , INC. IBALANCE TKA TIBIAL TRAY SIZE 7; PROSTHESIS, KNEE, PATELLOFEMORAOTIBAL, SEMI-CONTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX , INC. IBALANCE TKA TIBIAL TRAY SIZE 7; PROSTHESIS, KNEE, PATELLOFEMORAOTIBAL, SEMI-CONTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number IBALANCE TKA TIBIAL TRAY SIZE 7
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Swelling (2091)
Date of Event 06/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No further patient information was provided at the time of this report or made available in response to arthrex¿s follow-up requests.The device was not returned for evaluation and the contribution of the device to the reported event could not be determined.The information received regarding the event is not sufficient to determine the cause of the event.A dhr review for this catalog number, ar-503-ttte and lot number, 108761203 combination found no related information.
 
Event or Problem Description
It was initially reported by patient through the arthrex website, that the patient had total knee replacement procedure on both knees using arthrex products.(reference (b)(4)) patient provided the following information 2/13/18: initial surgery was (b)(6) 2013.Two years after surgery patient started to have swelling and pain in his knee.Surgeon had drained patient's knee three different times (around every 6 months), which brought him no relief.Surgeon informed patient that the compound (cement) used to adhere the implant was coming loose and he was most likely going to require a replacement.Patient stated his left knee is getting worse and his knee "slips out of place".Patient reported instability and pain in both knees that is so bad that sometimes he falls.Patient said his right knee is starting to have the same problems.When patient walks both knees shift.Patient stated he has pictures that show a "separation" of the implant.Arthrex requested that if a revision is performed, would he be willing to release the product to arthrex so that the implants could be inspected and so arthrex could report the findings back to patient.The following arthrex devices were implanted during the (b)(6) 2013 bi-lateral tka procedure: ar-503-ttth, lot 108761229, qty1; ar-503-ttth, lot 108761238, qty 1; ar-503-bh10, lot 113601221, qty 2; ar-503-pslj, lot 108761215, qty 1; ar-503-psrj, lot 108761219, qty 1; ar-504-psd0, lot 654468, qty 2.On 6/5/18 notice was received from patient that he is scheduled for left total knee revision surgery on (b)(6) 2018.Patient states revision is being performed due to left knee becoming loose and unstable.Original implant surgeon is scheduled to perform the revision procedure.Patient is requesting that his explanted devices be returned to arthrex for evaluation.Patient also stated that in the near future his right knee will need to be replaced for the same reasons.Additional information obtained (b)(6) 2018: arthrex has received confirmation that the planned revision procedure took place as scheduled on (b)(6) 2018.The following are the implants that were explanted on (b)(6) 2018: ar-503-ttth, lot 108761229, qty1; ar-503-bh10, lot 113601221, qty 1; ar-503-pslj, lot 108761215, qty 1; ar-504-psd0, lot 654468, qty 1.
 
Additional Manufacturer Narrative
The evaluation revealed that the returned device¿s critical mating features on surface a (surface that mates with the bearing) are within specification.The damages observed on the surfaces were most likely caused during the extraction of the implant.The information received regarding the event is not sufficient to determine the cause of the event.A dhr review for this catalog number ar-503-ttth and lot number 108761229 combination found no related information.
 
Event or Problem Description
It was initially reported by patient through the arthrex website, that the patient had total knee replacement procedure on both knees using arthrex products.(reference (b)(6)) patient provided the following information 2/13/18: initial surgery was (b)(6) 2013.Two years after surgery patient started to have swelling and pain in his knee.Surgeon had drained patient's knee three different times (around every 6 months), which brought him no relief.Surgeon informed patient that the compound (cement) used to adhere the implant was coming loose and he was most likely going to require a replacement.Patient stated his left knee is getting worse and his knee "slips out of place".Patient reported instability and pain in both knees that is so bad that sometimes he falls.Patient said his right knee is starting to have the same problems.When patient walks both knees shift.Patient stated he has pictures that show a "separation" of the implant.Arthrex requested that if a revision is performed, would he be willing to release the product to arthrex so that the implants could be inspected and so arthrex could report the findings back to patient.The following arthrex devices were implanted during the (b)(6) 2013 bi-lateral tka procedure: ar-503-ttth, lot 108761229, qty1; ar-503-ttth, lot 108761238, qty 1; ar-503-bh10, lot 113601221, qty 2; ar-503-pslj, lot 108761215, qty 1; ar-503-psrj, lot 108761219, qty 1; ar-504-psd0, lot 654468, qty 2.On 6/5/18 notice was received from patient that he is scheduled for left total knee revision surgery on (b)(6) 2018.Patient states revision is being performed due to left knee becoming loose and unstable.Original implant surgeon is scheduled to perform the revision procedure.Patient is requesting that his explanted devices be returned to arthrex for evaluation.Patient also stated that in the near future his right knee will need to be replaced for the same reasons.Additional information obtained 6/26/18: arthrex has received confirmation that the planned revision procedure took place as scheduled on (b)(6) 2018.The following are the implants that were explanted on (b)(6) 2018: ar-503-ttth, lot 108761229, qty1, ar-503-bh10, lot 113601221, qty 1, ar-503-pslj, lot 108761215, qty 1, ar-504-psd0, lot 654468, qty 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBALANCE TKA TIBIAL TRAY SIZE 7
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMORAOTIBAL, SEMI-CONTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
ARTHREX , INC.
1370 creekdside boulevard
naples FL 34108 1945
MDR Report Key7703770
Report Number1220246-2018-00556
Device Sequence Number789882
Product Code JWH
UDI-Device Identifier00888867106758
UDI-Public00888867106758
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K051726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Model NumberIBALANCE TKA TIBIAL TRAY SIZE 7
Device Catalogue NumberAR-503-TTTH
Device Lot Number108761229
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2018
Initial Date Received by Manufacturer 06/26/2018
Supplement Date Received by Manufacturer06/26/2018
Initial Report FDA Received Date07/19/2018
Supplement Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-