DEPUY MITEK HEALIX ADVANCE PEEK ANCHOR, 5.5 MM W/ORTHOCORD - 3 SUTURE; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
|
Back to Search Results |
|
| Catalog Number |
222307 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/22/2014
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported by the hospital that the surgeon was doing a rotator cuff repair and had already used two healix advance anchors.It was reported that the first two anchors went in great and had good purchase, but the third one pulled out when he went to pull the inserter out of the anchor.It was reported that the surgeon properly undid the sutures before pulling on the inserter to remove it and when he pulled the anchor came with the inserter.The surgeon had them open a new anchor, and to be safe, punched another hole in a new spot and the anchor held great.It was reported that the surgeon was not entirely convinced that it wasn¿t just a weak spot in the patients bone.It was not reported if there were any delays in the surgical procedure.But regardless did not want to pay for the anchor that pulled out.I was unable to attend this case, dr.(b)(6) and his pa reported this to me the day after the case.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|