Model Number 44010
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 07/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the patient was to have brain tumor surgery.During the operation, the reservoir was implanted; however, the ventricular tube broke and was immediately replaced with another device.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
The returned port met the requirements for leak testing.There was proteinaceous debris noted within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the catheter was returned in two pieces of lengths 13 cm and 10 cm respectively.The tear site appeared to be jagged.It is unknown how or when the damage occurred.The returned segments were patent and passed leak testing when tested individually.The instructions for use (ifu) that accompany the device caution that ¿low tear strength is a characteristic of most unreinforced silicone elastomer materials.Care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks or tears.¿ all ports and catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF VENTRICULAR PORT, 16 MM, SIDE INLET
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7704612
Report Number2021898-2018-00361
Device Sequence Number996201
Product Code LKG
UDI-Device Identifier00643169497108
UDI-Public00643169497108
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K874468
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model Number44010
Device Catalogue Number44010
Device Lot NumberE34897
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer08/28/2018
08/08/2018
Initial Report FDA Received Date07/19/2018
Supplement Report FDA Received Date09/06/2018
10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-