| Model Number |
44010 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/17/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2018, the patient was to have brain tumor surgery.During the operation, the reservoir was implanted; however, the ventricular tube broke and was immediately replaced with another device.The patient's status at the time of the report was alive-no injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned port met the requirements for leak testing.There was proteinaceous debris noted within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the catheter was returned in two pieces of lengths 13 cm and 10 cm respectively.The tear site appeared to be jagged.It is unknown how or when the damage occurred.The returned segments were patent and passed leak testing when tested individually.The instructions for use (ifu) that accompany the device caution that ¿low tear strength is a characteristic of most unreinforced silicone elastomer materials.Care must be taken with the handling and placement of the silicone elastomer catheter tubing to avoid cuts, nicks or tears.¿ all ports and catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|