Model Number 5515-F-602
Medical Device Problem Codes Unstable (1667); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Joint Laxity (4526)
Date of Event 06/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that patient's right knee was revised due to a surgical error resulting in instability.Rep reported that the gaps were not balanced and that the knee was stable in extension but loose in flexion.Patient was revised to a triathlon ts construct with a larger size 7 femur.Rep reported that x-rays, medical records, and further information are not available due to hospital policy.
 
Event or Problem Description
It was reported that patient's right knee was revised due to a surgical error resulting in instability.Rep reported that the gaps were not balanced and that the knee was stable in extension but loose in flexion.Patient was revised to a triathlon ts construct with a larger size 7 femur.Rep reported that x-rays, medical records, and further information are not available due to hospital policy.
 
Additional Manufacturer Narrative
Corrected data: it was discovered that the selection for b2 was inadvertently missed during the submission of the initial report.B2 has been updated with the appropriate information.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS FEM COMPONENT, CEMENTED
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7704729
Report Number0002249697-2018-02251
Device Sequence Number16805260
Product Code MBH
UDI-Device Identifier07613327050424
UDI-Public07613327050424
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5515-F-602
Device Catalogue Number5515-F-602
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/21/2018
Supplement Date Received by Manufacturer11/23/2020
Initial Report FDA Received Date07/19/2018
Supplement Report FDA Received Date11/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight96
-
-