| Model Number |
MMT-751NAP |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call blank display anomaly on the insulin pump.Customer¿s blood glucose level was 262 mg/dl.Troubleshooting for blank display was performed.Customer advised they changed the battery and the battery cap however the insulin pump remained blank.Customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the operating currents, self test and displacement test.No blank display anomaly noted during test.
|
| |
|