| Model Number |
SL600HD |
| Medical Device Problem Code |
Component Misassembled (4004)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
05/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This type of product performance issue has not been reported previously.During the on-site incident investigation, several screws that connect the chassis motor to the frame were found to be loose or missing.The unit was sent to harmar for further investigation in may where it was determined that the screws were loose or missing due to the lack of loctite threadlocker application on the screws.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 the user got on the stair lift chair at the bottom step and was riding upwards to the top landing when the stair lift made a popping sound and the stair lift seat/chassis descended back down the staircase to the end of the rail.The user stepped off the stair lift as it descended and sustained minor pain in her leg and a small cut on her finger.
|
| |
|