COVIDIEN TRAILBLAZER SUPPORT CATHETER; CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number SC-018-150
Health Effect - Clinical Codes Foreign Body In Patient (2687); Patient Problem/Medical Problem (2688)
Date of Event 06/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The physician was attempting to use a trailblazer to treat a lesion in the left superficial femoral artery (sfa).It was reported that the procedure was difficult due to the severity of arterial calcification.When passing the device through the calcification, the micro catheter on the wire was detached on the proximal end.The detached end of the micro catheter was retrieved using a non-mdt snare capture device.The wire and the catheter were able to be removed and replaced.Nothing was left in the body and the patient did not sustain any harm as a result.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAILBLAZER SUPPORT CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COVIDIEN
4600 nathan lane north
plymouth MN 55442
MDR Report Key7706209
Report Number2183870-2018-00389
Device Sequence Number1404867
Product Code DQY
UDI-Device Identifier00821684073926
UDI-Public00821684073926
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K092299
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2020
Device Catalogue NumberSC-018-150
Device Lot NumberA554449
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/22/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date07/20/2018
Supplement Report FDA Received Date10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-