• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Catalog Number 07976887190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred in (b)(6).
 
Event or Problem Description
The customer questioned results for 1 patient tested for elecsys ft4 ii (ft4 ii), tsh, anti-tpo and anti-tshr on a cobas e 801 module.Refer to the associated medwatches with the following patient identifiers: (b)(6).Between (b)(6) 2017 and (b)(6) 2018 the patient had increased ft4 ii results (29-52 pmol/l) and low tsh results (0.04 - 0.17 mie/l) in 14 different samples from the patient when tested on the e801 module.Anti-tpo results were > 600 ie/ml in 6 different samples from the patient when tested on the e801 module.The patient previously had negative anti-tshr results in 2 different samples when tested by the (b)46) method, however, when the customer changed to the roche method in (b)(6) 2017, the patient had positive anti-tshr in 7 different samples when tested on the e801 module.The actual results were not provided.The roche results were reported outside of the laboratory where the patient was diagnosed with hyperthyroidism and treatment was started with neo-mercazole in (b)(6) 2017.The customer suspects an interference in the patient sample as the patient's clinical symptoms do not agree with the results.The customer sent 3 patient samples obtained in (b)(6) 2018 and (b)(6) 2018 for testing by the abbott architect method.Discrepant results were identified for ft4 ii and tsh between the e801 module and the abbott method.Refer to attached data for the patient comparison results.After the patient samples were tested by the abbott architect method, neo-mercazole treatment was stopped.There was no allegation that an adverse event occurred.The e801 module serial number was (b)(4).
 
Additional Manufacturer Narrative
One sample from the patient was submitted for investigation.An interfering factor was not identified.A product problem was not found.Investigations are ongoing.
 
Additional Manufacturer Narrative
It was clarified that the ft4 ii results from the initial report were actually elecsys ft4 iii (ft4 iii) results.Upon further investigation of the patient sample, an interferent against a component of the reagent was confirmed.This interference is covered in product labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Common Device Name
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7706293
Report Number1823260-2018-02414
Device Sequence Number1365607
Product Code CEC
UDI-Device Identifier04015630939725
UDI-Public04015630939725
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
K131244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/04/2018
Supplement Date Received by Manufacturer07/04/2018
07/04/2018
Initial Report FDA Received Date07/20/2018
Supplement Report FDA Received Date08/14/2018
09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LIPITOR; NEO-MERCAZOLE; PREDNISOLONE
Patient Age70 YR
-
-