Model Number H7493952816270
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).The complaint device was not received for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is device not returned as the complaint did not include returned device review and lacked the objective evidence or descriptive conditions of the event required to determine what could have contributed to the event.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.A 2.75x16mm promus premier¿ drug-eluting stent was advanced to treat the lesion.However, while inserting the device into the guide, the shaft broke midway from the hub.The device was removed normally.The procedure was completed with another of the same device.No patient complications were reported and the patient status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7706385
Report Number2134265-2018-06342
Device Sequence Number1588911
Product Code NIQ
UDI-Device Identifier08714729844723
UDI-Public08714729844723
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/27/2019
Device Model NumberH7493952816270
Device Catalogue Number39528-1627
Device Lot Number22087659
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2018
Initial Report FDA Received Date07/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No