| Catalog Number |
01.12.024 |
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 20 july 2018; lot 176099: (b)(4) items manufactured and released on 18 january 2018.Expiration date: 2023-01-10; no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
The surgeon templated for a size 4/5 quadra std stem and broached up to a size 4 using a std lateral approach surgical technique.He performed a trial reduction and then double checked the torsional and lineal stability prior to calling for the definitive stem.Upon implantation of the size 4 stem he found it to subside approximately 1 cm.He then removed the stem and broached up to a size 5 and ultimately implanted that stem which sat 3mm deeper than broached.The surgeon felt that it was stable and accepted the position.
|
| |
|