• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK SPHERE TIBIAL TRAY FIXED CEMENTED SIZE T4-I3 R

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK SPHERE TIBIAL TRAY FIXED CEMENTED SIZE T4-I3 R Back to Search Results
Catalog Number 02.12.T4I3R
Health Effect - Clinical Code Pain (1994)
Date of Event 06/22/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 20 july 2018: lot 153123: (b)(4) items manufactured and released on 12 june 2015.Expiration date: 30.04.2020 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Other device involved: reference 02.12.0003r (k121416) femoral component sphere cemented # 3 r lot.161657 lot 161657: (b)(4) items manufactured and released on 26 april 2016.Expiration date: 11.04.2021 no anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs director on 20 july 2018: 1,5 years after primary cemented tka the patient needed revision.The radiographs supplied clearly show that the position of the components is most likely the cause of the problem.The alignment is not adequate and the tibial tray does not cover properly the tibial metaphysis.We have only one set of xray therefore we cannot tell if this position is the result of component migration or if it was a surgical decision.We have no data reporting any particular surgical strategy.There is no reason to suspect malfunctioning devices.
 
Event or Problem Description
Due to malpositioning of implants (primarily femoral component, but also tibial component), the patient suffered residual pain and needed a revision 1 year and 8 months after revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK SPHERE TIBIAL TRAY FIXED CEMENTED SIZE T4-I3 R
Common Device Name
TIBIAL TRAY FIXED CEMENTED
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
MDR Report Key7706402
Report Number3005180920-2018-00531
Device Sequence Number16326714
Product Code JWH
UDI-Device Identifier07630030827167
UDI-Public07630030827167
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K121416
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue Number02.12.T4I3R
Device Lot Number153123
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2018
Initial Report FDA Received Date07/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient Weight76
-
-