• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD VERSAPORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number ONB5STF2C
Medical Device Problem Codes Detachment of Device or Device Component (2907); Packaging Problem (3007)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during laparoscopic cholecystectomy, the device had a damaged seal.When inserting the device into sleeve, a small piece of plastic blue valve shaved off and had to be retrieved from patient.They opened another device to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of two devices.The visual inspection noted: the circular seals, trocars and envelope seals appeared intact.The obturator was received and appeared intact.A functional evaluation found that the devices passed an air leak test.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7706415
Report Number9612501-2018-01334
Device Sequence Number1269782
Product Code GCJ
UDI-Device Identifier10884521657656
UDI-Public10884521657656
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K112349
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model NumberONB5STF2C
Device Catalogue NumberONB5STF2C
Device Lot NumberJ8B1270JX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Initial Date Received by Manufacturer 07/05/2018
Supplement Date Received by Manufacturer08/08/2018
09/14/2018
Initial Report FDA Received Date07/20/2018
Supplement Report FDA Received Date10/03/2018
10/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-