Model Number MICL12.6
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); No Code Available (3191)
Date of Event 06/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Work order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted 12.6mm micl12.6, -16.0 diopter implantable collamer lens into the patient's left eye (os) on (b)(6) 2018.On (b)(6) 2018 the lens was removed by a surgeon that was on call due to observation of excessive vault, significant reduction of irido-corneal angles, pupil block and angle closure with elevated iop (intraocular pressure).Additional surgical intervention (enlargement and addition of new pi, yag) were performed.On (b)(6) 2018 an exchanged for a shorter length lens was performed.This exchange resolve the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7706773
Report Number2023826-2018-01067
Device Sequence Number1588337
Product Code MTA
UDI-Device Identifier00841542103145
UDI-Public00841542103145
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model NumberMICL12.6
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2018
Initial Report FDA Received Date07/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: AQ-FP, LOT# 1313619; INJECTOR MODEL: MSI-PF, LOT# 1401906; OVD TYPE: OCUCOAT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
-
-