If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: evaluation statement: the complaint device is not being returned, therefore is unavailable for a physical evaluation.A batch record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system revealed no other complaints of any kind for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: the batch review 3738139 showed that the devices were conformed to in process and finished goods specifications when released to stock the 10 dec.2013.Expiration date: 2016-10 (according to retained labels).No nc was opened.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
|
Surgeon has now used the healix advance knotless anchor on eight patients (17 anchors).When he has good mobility in the tendon to be repaired, the anchor works quite well.However, on occasion he needs to mobilize the tendon under tension in order to complete the repair.He complains that the anchor does not perform well in this situation.As the sutures exit out the bottom of the anchor, when he tries to tension them it causes the anchor to be pulled out of the prepared hole despite him putting significant downward pressure on the insertion handle.In this procedure, this caused the distal tip of one anchor to deform and it then had to be removed.Additionally, when he had six limbs of sutures, he felt that they obstructed his vision and hampered docking of the anchor tip into the prepared hole.Surgeon suggests that the manufacturer look at putting a window near the distal tip of the anchor, just below the first thread, so that the tip can be docked into the hole securely and he can then tension the sutures.Additionally, he does not use a clear cannula in these cases, so the white docking tab cannot be used as designed because it tends to trampoline off the patients skin.Another suggestion is to have the dock fixed at the proximal end of the insertion shaft.Case completed with same like product.A 5-min delay.No ae to patient.
|