• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; SCREW Back to Search Results
Model Number 25-092-08-91
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/28/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Reference exemption number e2017029.Multiple mdr report were filed for this event, please see associated report: mdr 9610905-2018-00141.
 
Event or Problem Description
It was reported while placing the mmf screw the head broke off.
 
Additional Manufacturer Narrative
An investigation was performed using a stereo microscope which revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.A review of the device history records was not possible due to no lot number identified.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau 78570
GM  78570
MDR Report Key7706945
Report Number9610905-2018-00140
Device Sequence Number1365623
Product Code DZL
UDI-Device Identifier00888118037206
UDI-Public(01)00888118037206
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K042573
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
Report Date (Section F) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-092-08-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/28/2018
Event Location Hospital
Date Report to Manufacturer06/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/24/2018
Initial Report FDA Received Date07/20/2018
Supplement Report FDA Received Date09/14/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-