• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL12.6
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Lens was returned dry, stuck to lens vial.Bss was added to remove lens from vial.Visual inspection found a torn haptic and clear surgical residue on the lens surface.No similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The surgeon stated that a 12.6mm micl12.6, -14.0 diopter, implantable collamer lens was injected into the patient's left eye (os) and it flipped over in the ac (anterior chamber).The lens was removed from the anterior chamber and another lens was implanted during the same surgery.The lens tore during removal but there was no patient injury.This exchange resolved the problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7707568
Report Number2023826-2018-01069
Device Sequence Number11160209
Product Code MTA
UDI-Device Identifier00841542103107
UDI-Public00841542103107
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2019
Device Model NumberMICL12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2018
Initial Report FDA Received Date07/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
-
-