Exemption number: e2013009.Berlin heart inc.(importer number: (b)(4)) is submitting the report on behalf of berlin heart gmbh (manufacturer).The excor driving tube, lot number 00060195, was used from 2018-03-12 to 2018-06-26 (106 days).During investigation, a break was detected at the transition to the pump connector.At this region of the driving tube, there is a transition from smaller diameter to larger diameter.This transition area of the driving tube is where the most stress is applied by external forces when in use by a patient.This driving tube lot was manufactured after implementation of changes by the manufacturer in 2015.Upon further evaluation of the driving tube, small cracks were detected on the surface of the tubing close to the break, which indicate that a high degree of movement had occurred at this region.According to the (b)(4) distributor, the driving tube was stressed due to the sitting position of the patient.This is the second case involving a similar failure scenario, observed with this patient.The cause of the defect was most likely due to the positioning of the driving tube in combination with the movement patterns of this patient.The exact cause of the defect cannot be detected in this case.According to the "important safety information" in the instructions for use, the user is cautioned to ensure that there are no kinks in the blood pumps, cannulae and driving tubes.Otherwise, support could be inadequate.Additionally, users are reminded to ensure that the blood pumps, cannulae and driving tubes are never exposed to tension, torsion, pressure or strain.Otherwise, support could be inadequate or tension may cause damage to the transcutaneous exit site.
|