• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ECHELON 45 ENDOPATH STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ECHELON 45 ENDOPATH STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number EC45
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The lot/batch was not provided; therefore, the manufacturing records could not be reviewed.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested and the following was obtained: can you please confirm the product code for the complaint device? is it ec45? yes.Or, is it ec45a? attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Did any piece of the reload fall into the patient? if yes, how was it retrieved? did this alter the procedure in any way?.
 
Event or Problem Description
It was reported that during an unknown procedure, the knife cuts into the magazine plastic parts from the magazine found in the instrument mouth.No harm to the patient.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information requested and received: did any piece of the reload fall into the patient? no.If yes, how was it retrieved? did this alter the procedure in any way? no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON 45 ENDOPATH STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7707778
Report Number3005075853-2018-11465
Device Sequence Number19086529
Product Code GDW
UDI-Device Identifier20705036001666
UDI-Public20705036001666
Combination Product (Y/N)N
PMA/510(K) Number
K051002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEC45
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/28/2018
Supplement Date Received by Manufacturer08/31/2018
Initial Report FDA Received Date07/20/2018
Supplement Report FDA Received Date09/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-