| Model Number |
5177801400 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/26/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, tubing broken in both cylinders near pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Quality was unsuccessful in securing the explanted prosthesis for evaluation.Without the benefit of analyzing the explant, quality cannot confirm any observations and cannot comment on the condition of the prosthesis.If the explanted device becomes available, or additional information is received, quality will re-evaluate this complaint in accordance to procedures.Quality reviewed the manufacturing records for this device and confirmed that there were no discrepancies that would have contributed to this complaint and verified that the devices from this lot met all specifications prior to release.A review of the complaint history database revealed no significant trends for lot 1680610.Review of nonconforming reports revealed no nonconformance's for this lot.No capas are associated with this lot.Management routinely reviews events such as this and monitors complaint levels.Additionally, events of this type are captured in the product risk documentation.Based on this information, no further corrective action is required at this time.
|
| |
|