| Model Number |
TN, LH780 ANALYTICAL STATION |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Test Result (2695)
|
| Date of Event |
06/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2018 the field service engineer was at the customer site and observed bubbles coming in from the fitting on the clenz input reservoir causing low mchc results.On (b)(6) 2018 the fse returned because the customer reported having low mchc results again.This visit the fse observed that in addition to the low mchc, the coulter lh 780 hematology analyzer was also generating high rbc and platelet results intermittently.The fse replaced the aspiration pump, cleaned and greased blood sampling valve (bsv) mechanism, rerouted the bsv tubing to remove possible stress points caused by the geared probe wipe assembly.Patient information was not provided by the customer.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported having low mchc for patient samples on their coulter lh 780 hematology analyzer.Erroneous results were not reported outside of the laboratory.There was no change or effect to patient treatment as a result of this event.
|
| |
|