If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.The tip of the inserter is broken confirming this complaint.From past investigations, this type of failure can be attributed to exerting too much pressure on the device while using or dropping the device on the ground.Other than this possibility of occurrence, we cannot determine a root cause for this issue.It is unknown at what point this failure occurred.Furthermore, no lot numbers were supplied which precludes conducting a batch history review or a lot specific search in the complaints handling system.At this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4) - incomplete.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|