| Model Number |
60726807 |
| Medical Device Problem Codes |
Smoking (1585); Defective Device (2588); Electrical Overstress (2924); Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the limited information, we are unable to determine the root cause of the alleged incident; however, the equipment is scheduled to return to pittsburgh for a full evaluation.Once the evaluation is completed, a follow-up report will be submitted.This information does not constitute an admission that the device, the company, or its employees caused or contributed to a reportable event.
|
| |
|
Event or Problem Description
|
|
A bayer representative had reported seeing smoke coming out of the stellant ct injector base unit.The injector was powered down and unplugged from the wall.No injury or adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Bayer service replaced the base unit and confirmed system functionality.Bayer product analysis received and examined the returned base unit.Visual examination confirmed that there were degraded components on the 15v power supply which are consistent with a typical electronic fault.The power supply is centrally located within a metal electronic chassis of the base unit of the stellant injector and is well isolated from the patient and user.Therefore, there is no injury or property damage risk if the fault were to recur.
|
| |
|