• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICAN REPUBLIC INC. EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICAN REPUBLIC INC. EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 471954
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 02/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).Ten of the received samples from lot 0061555072 were visually examined per specification with no efects noted.Those ten samples were occlusion and leakage tested per specification with passing results.The returned unused sets did not confirm the reported defect of leakage.Review of the discrepancy management system database performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.We will maintain this report for further reference and continue to monitor other reports for similar occurrences.
 
Event or Problem Description
As reported by user facility: while administering the electrolyte solution to a nicu patient, the nurse found that the female luer was cracked causing about 5ml of blood to leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
Manufacturer (Section G)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR  
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key7710344
Report Number9614279-2018-00126
Device Sequence Number11480596
Product Code FPA
UDI-Device Identifier04022495770103
UDI-Public(01)04022495770103
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K942988
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number471954
Device Catalogue Number471954
Device Lot Number0061555072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/06/2018
Supplement Date Received by Manufacturer03/06/2018
Initial Report FDA Received Date07/23/2018
Supplement Report FDA Received Date07/23/2018
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/23/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-