• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUETRACK; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code Polydipsia (2604)
Date of Event 07/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Returned meter evaluated with no defect found.Test strip vial returned with expired test strips - unable to test.Most likely underlying root cause: mlc-12 product expired test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call in order to ensure the customer's condition since the initial call - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for hi blood glucose test results.Nurse is calling on behalf of the customer.The customer is concerned with tests results from results obtained of hi.The customer's expected fasting blood glucose test result range is 200 - 300 mg/dl.At the time of the call the customer reported feeling cold and thirsty.Nurse was with customer and stated she would call physician.During the call on (b)(6) 2018, a back to back blood test was not performed by the customer.The product is stored according to specification in the living room.The test strip lot manufacturer's expiration date is 04/30/2018; strips are expired and nurse was aware.Customer did not remember the open vial date.Customer did not have another vial of test strips.The meter memory was reviewed for previous test result history: (b)(6).The nurse says she is calling because she says that meter is old and believes that it is defective.Strips are expired and nurse is aware.No other strips are available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7710482
Report Number1000113657-2018-00780
Device Sequence Number19214433
Product Code NBW
UDI-Device Identifier00096295129489
UDI-Public(01)00096295129489
Combination Product (Y/N)N
Initial Reporter StateWI
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 07/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2018
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRS4814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/02/2018
Initial Report FDA Received Date07/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY
-
-